Summary
Type: SPE
Application #: 305070
Catalog#: 730015 - SPE Cartridges, Chromabond, C18 ec, 6ml, PP, 1000mg, 30/pk
Detection:
Conditions: Automated MPS-SPE method
Column conditioning: 10 ml methanol, then 10 ml water
Sample aspiration: 5 ml sample onto MPS-C18 ec column
Washing: 5 ml water
Elution: Analytes are eluted with 5 ml acetonitrile/water (600 µl/min with a reproducible pressure)
Evaporation: 50 °C, for 6 min
Reconstitution: 5 ml acetonitrile/5 mM acetic acid (30:70)
Authors: Helle, N.; Lerch, O; Baden, M.
Source: GIT Spezial, Separation 1/2009, 36-37